Abstract

We measured Raman spectra of N2 fluids obtained at 0.1-25MPa at room temperature. The 14N15N peak in the Raman spectrum of a low-pressure N2 fluid is difficult to detect because of the prevalence of a group of peaks attributed to rotational vibration of 14N2. The Raman peaks of 14N15N and 14N2 of N2 fluid at 25MPa were acquired at various exposure times. The mean values and standard deviations of the peak height ratios and peak area ones of 14N15N and 14N2 were examined for each time. The standard deviations of the peak height ratios and peak area ones were 2.2% and 1.9%, respectively, for 20 spectra acquired with peak height of 1 million counts of 14N2. The uncertainties are about two times higher than those predicted from the noise of a CCD. Improvement of the pixel resolution can enhance the precision of the nitrogen isotope ratios by Raman spectroscopy.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.