Abstract

A precision electromagnet generating a magnetic field with an induction ranging from 0.05 to 0.50 T is designed, manufactured, and studied. It is intended for a magnetic resonance mass spectrometer with a rated resolution of about 106. The magnetic field inhomogeneity on a circular orbit with a diameter of 400 mm along which the ion beam moves is no more than ±1 × 10−5 of induction B0 at the center of the magnetic gap. At any point of the orbit, the magnetic field is kept constant with an accuracy of higher than 10−6 for several minutes, which is sufficient to record mass spectra.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.