Abstract

To have confidence in force measurements made with atomic force microscopes (AFMs), the spring constant of the AFM cantilevers should be known with good precision and accuracy, a topic not yet thoroughly treated in the literature. In this study, we compared the stiffnesses of uncoated tipless uniform rectangular silicon cantilevers among thermal, loading, and geometric calibration methods; loading was done against an artifact from the National Institute of Standards and Technology (NIST). The artifact was calibrated at NIST using forces that were traceable to the International System of units. The precision and accuracy of the thermal method were found to be 5% and 10%, respectively. Force measurements taken with different cantilevers can now be meaningfully compared.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.