Abstract

Semi-active robot-assisted total hip arthroplasty (THA) has two options to measure the leg length discrepancy (LLD) and combined offset (CO), the 'enhanced' femoral workflow and the so-called 'express' simplified workflow. The purpose of this study was to determine the precision and accuracy of intraoperative LLD and CO measurement with express workflow robotic THA. Between February 2018 and December 2019, 30 patients underwent an express workflow robot-assisted primary THA for intraoperative LLD and CO measurements. Postoperative radiographs were used for LLD and CO measurement. In order to examine the accuracy of the robotic system assessment, the absolute difference between the robotic assessments and radiographic evaluations was calculated. Intraoperative robotic measurements reported a mean error of 0.2-0.6 mm for each registration, with no significant difference between them (p = 0.311). The average absolute discrepancies between the robotic and radiographic assessments in the LLD and CO measurements were 1.3 ± 1.5 mm (p = 0.17) and 1.1 ± 0.9 mm (p = 0.11), respectively, while the Pearson's correlation coefficients were 0.69 and 0.71. An external marker without a femoral array inserted into a screw positioned in the greater trochanter would be an easier and faster method to measure LLD and CO. Our study showed that the measured values of LLD and CO obtained by intraoperative express workflow robot-assisted THA system were precise and accurate.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.