Abstract

Point-of-care (POC) glucometry is commonly used in horses; however, measurement error with this method when analysing hypoglycaemic samples (<4 mmol/L) is unknown. The objective of this study was to determine the precision and accuracy of glucometry in hypoglycaemic horses in comparison to a laboratory method of glucose measurement (LAB). Repeatability coefficients were 0.47 mmol/L for POC and 0.09 mmol/L for LAB, and coefficients of variation were 10 % and 2.11 %, for the POC and LAB methods, respectively. Systemic bias with the POC method was present, with a mean bias of −0.26 mmol/L (95 % limits of agreement: −0.88 – 0.37) in comparison to LAB, and <70% of measurements were within 20 % of paired LAB results. Prior to use of glucometers, assessment of the diagnostic performance of the equipment is necessary, including determination of acceptable criteria and reference ranges for hypoglycaemic samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.