Abstract

Three widely used methods to calculate the response functions for NaI(Tl) detectors were investigated. The methods were Berger–Seltzer's method (Nucl. Instrum. Methods 104 (1972) 317), general Monte Carlo (MC) programs, such as EGS4 (The EGS4 code system, Stanford Linear Accelerator Center, 1985) and MCNP4B (MCNP—a general Monte Carlo N-particle transport code, Los Alamos National Laboratory Report, LA-12625-M, 1997), and special MC programs. The pulse height spectra in a 3″×3″ NaI(Tl) detector due to several gamma-ray sources have been measured to verify the calculated results of these methods. The energies of the sources ranged from 0.4118 to 7.11 MeV. The spectra generated by Berger–Seltzer's method and the general MC programs did not agree well with the experimental data. PETRANS 1.0, the special MC program developed in house, was fairly accurate since it also considered the scintillation efficiency and the single escape peak shift.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.