Abstract

We present a method of precise measurements of spectral line profiles that was elaborated and tested on the Main Stellar Spectrograph (MSS) of the 6-m BTA telescope. The method allows discarding corrections for flat field. To take account of inhomogeneity of the CCD chip sensitivity, two spectra are recorded, shifted against each other along the dispersion by one column. Recurrent ratios between the photo responses of the CCD pixels let us determine their relative sensitivity and correct the line profiles. To eliminate various instability effects, the spectra are registered with the use of a cyclic shift procedure along the dispersion, and a synchronous electronic image transfer across the spectrum dispersion (along the CCD matrix columns).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.