Abstract

High resolution inductively coupled plasma mass spectrometry (HR-ICP-MS) was used to measure Pb isotope ratios in standard solutions and Pb mineral digests. The RSD values obtained for208Pb/204Pb, 207Pb/204Pb, 206Pb/204Pb, 208Pb/206Pb and 207Pb/206Pb were 0.13, 0.11, 0.11, 0.046 and 0.048%, respectively (values as 1σ). These values were obtained from 30 analyses of three different standard sample types (multi-element standard, NIST SRM 981 and a Broken Hill galena digest). Based on 39 analyses of 11 galena samples from different locations around Australia, the difference between HR-ICP-MS and conventional TIMS values for 208Pb/204Pb, 207Pb/204Pb, 206Pb/204Pb, 208Pb/206Pb and 207Pb/206Pb ratios was generally better than±0.2%. This paper outlines a very simple and rapid analytical method for the measurement of Pb isotope ratios, and is one of the first studies to use HR-ICP-MS to measure Pb isotope ratios in galena and galena-bearing ores.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.