Abstract

ABSTRACT Background This article presents a strategy that a Drug Delivery Device Developer (DDDD) has adopted to support Abbreviated New Drug Application (ANDA) submissions of drug-device combination products. As per the related FDA guidance, a threshold analysis should be compiled. If ‘other differences’ between the Reference Listed Drug (RLD) and the generic drug devices are identified, a Comparative Use Human Factors (CUHF) study may be requested. Methods The DDDD performed task analysis and physical comparison to assess the pen injector design differences. Then, a formative CUHF study with 25 participants simulating injections using both RLD and the generic pen injectors was conducted. Results After each participant completed four simulated injections, similar type and rates of use error between the RLD (0.70) and generic (0.68) pen injectors were observed. Conclusion DDDDs can support pharmaceutical companies in the ANDA submission strategy of their drug-device combination product by initiating comparative task analysis and physical comparison of the device as inputs for the threshold analysis. If ‘other differences’ are identified, a formative CUHF study can be performed. As shown in our case study, this approach can be leveraged to support the sample size calculation and non-inferiority margin determination for a CUHF study with the final combination product.