Abstract

We have developed a very low noise preamplifier that meets the low-cost, high-density requirements for use with superconducting tunnel junction (STJ) arrays. The design allows the STJs' bias points to be digitally set at the μV level. We then trace out noise versus bias curves under computer control and, by showing the correlation between this noise and detector energy resolution, establish the ability to automatically select optimum STJ operating points. Working with a 75 μm × 75 μm STJ with a Ta x-ray absorber, we achieved better than 7 eV FWHM at the 525 eV O-K line with about 2 eV coming from the preamplifier.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.