Abstract

Reports on the accuracy of magnetic resonance angiography (MRA) and magnetic resonance venography (MRV) in evaluating living donor renovasculature employ few patients or omit the consequences of inaccurate scans. We retrospectively compared intraoperative findings to MRA/MRV scans in 146 donor-recipient pairs. For detecting accessory arteries and early branching, MRA sensitivity was 57.6%, specificity 96.5%, false positive rate 3.5%, false negative rate 42.4%, positive predictive value 82.6%, negative predictive value 88.6% and overall accuracy 87.7%. By excluding clinically inconsequential accessory arteries, MRA sensitivity rose to 73.1%, specificity to 96.7% and overall accuracy to 92.5%. For MRVs, sensitivity was 56.2%, specificity 99%, false positive rate 1%, false negative rate 43.8%, positive predictive value 90%, negative predictive value 94.8% and accuracy 94.5%. Inaccurate scans were associated with prolonged donor and recipient operations and more frequently reconstructed arteries, but did not affect clinical outcomes. Because most missed accessory arteries are inconsequential, MRA is a useful, less invasive method for defining donor renovascular anatomy.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.