Abstract

A bismuth pre-concentration and determination system implemented with inductively coupled plasma atomic emission spectrometry (ICP-AES) associated with a flow-injection (FI) method with an ultrasonic nebulization system was studied. For the retention and on-line pre-concentration of bismuth, quinolin-8-ol and Amberlite XAD-7 were used at pH 5.0. The bismuth complex was removed from the micro-column with nitric acid. The detection limit for the pre-concentration of 100 ml of aqueous solution was 0.03 ng ml –1 with a relative standard deviation (RSD) of 2.5%, calculated with the peak heights obtained. The RSD for the human urine samples were between 1 and 5%. This reflects the cumulative imprecision of all the process. The calibration graph using the pre-concentration system for bismuth was linear with a correlation coefficient of 0.9991 at levels near the detection limits up to at least 100 ng ml –1 . The method was succesfully applied to the determination of bismuth in human urine samples digested previously with a mixture of H 2 O 2 and HNO 3 .

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.