Abstract
Pre-concentration and determination of Zn (II) across a ligand assisted pseudo stirbar hollow fiber solid/liquid phase microextraction method in water samples has been investigated. All detections were carried out by differential pulse anodic stripping voltammetry (DPASV). The method involves microextraction and pre-concentration of Zn (II) on the pseudo stir bar hollow fiber. Then desorption has been done using suitable solvent containing suitable ligand as complexing agent. The optimized conditions were obtained. The relationship between the peak current and concentration was linear over the range of 0.05–500ngmL−1. The limit of detection was 0.015ngmL−1. Under the optimized conditions, the pre-concentration factor is 5140. The applicability of the developed technique was evaluated by application to spiked, environmental water samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.