Abstract

Peptides larger than 3–4 kDa, such as neuropeptide Y (NPY), orexin-B, and alpha-MSH, have practical issues that arise when conducting direct and sensitive quantitative liquid chromatography (LC) orbitrap-FT mass spectrometry (MS) due to their adsorption and low ionization efficiency, especially in standard solutions. A mixing solvent consisting of 0.5% trifluoroacetic acid (TFA) and 35–50% aq. acetonitrile was developed as the standard NPY for creating calibration curves, as well as a matrix to block the experimental tube surface to minimize adsorption. The mixture matrix effectively blocked non-specific adsorption of the standard peptides with tryptic digested bovine serum albumin (BSA) (small fragment peptides) and orexin-B (a large chain peptide). A sample containing 1 : 100 peptide:water was detected in the developed sample solution. Finally, 2 to 1,000 fmol/μL NPY could be analyzed quantitatively and reproducibly using conventional LC-MS. Parameters of the calibration curves, such as X-intercept, Bias (%), and relative standard deviation (RSD), were adjusted to optimize the sample solutions and the sensitive and quantitative LC-MS analyses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.