Abstract

The practical problems in the implementation of the spatially resolved spectroscopy (SPARS) sequence have been addressed. Phantom studies have shown that the slice profile is distorted due to the spurious echoes and the phase shift induced by the settling gradient fields. Studies indicate that the spurious echoes arise because of the imperfect 180° refocusing pulse within the localization pulse train. It has been demonstrated that these echoes can be eliminated by the application of two orthogonal gradients around the 180° pulse. The phase shift introduced by the settling gradient fields has been compensated by phase shifting the selective 90° pulse with respect to the 180° refocusing pulse. It is shown that the spurious echoes do not significantly compromise the quality of the localized spectrum.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.