Abstract

Within near-field goniophotometry, measurement results of both an imaging luminance measurement device and a photometer detector are combined to generate the luminous intensity distribution of a light source. The simultaneous use of these two detectors may engender incorrect measurement results, due to their difference in dynamic range. In this paper, near-field and far-field based luminous intensity distribution measurements of two luminaires are presented, in order to exemplify the problem. Results demonstrate that the distributions obtained from near-field measurements may deviate from the correct intensity distribution, by an amount of up to 16% of the total luminous flux of the luminaire. A method to check for the correctness of the luminous intensity distribution from the near-field measurement, the so-called sanity check, is discussed. To conclude, some possible solutions to eliminate the dynamic range mismatch induced errors are treated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.