Abstract
Possibilities and practical aspects of implementation of splitless injection of larger volumes for fast GC purposes utilizing narrow-bore column, hydrogen as carrier gas, fast temperature programming under programmed flow conditions and commercial instrumentation were searched. As a model sample semivolatile compounds of a broad range of volatility and polarity (7 n-alkanes and 19 pesticides) were chosen. Peak shapes, peak broadening and peak areas and its repeatability were evaluated under various experimental set-ups (liner/injection technique combinations). Various factors, such as liner design, injection technique, retention gap length, compound volatility and polarity, the solvent used, initial oven temperature influenced compound focusation and/or maximal injection volume. Combination of analytical column (CP-Sil 13 CB 25 m long, 0.15 mm i.d., film thickness 0.4 μm) with normal-bore retention gap (1 m long, 0.32 mm i.d.) allowed maximal injection volume 8 μl for 4 mm i.d. liner used without any peak distortion when solvent recondensation in the retention gap was employed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.