Abstract

High resolution magic angle spinning (HRMAS) has become an extremely versatile tool to study heterogeneous systems. HRMAS relies on magic angle spinning of the sample to average out to zero magnetic susceptibility differences in the sample and to obtain resonance linewidths approaching those of liquid state NMR. Shimming such samples therefore becomes an important issue. By analyzing the different sources of magnetic field perturbations present in a sample under MAS conditions, we propose a simple protocol to obtain optimum shim settings in HRMAS. In the case of aqueous samples, we show that the lock level cannot be used as a reliable indicator of the quality of the shims at high spinning speeds. This effect is explained by the presence of temperature gradients imparted by the sample rotation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.