Abstract

Incurred sample reanalysis (ISR) is an important step in assuring the quality of an LC-MS/MS bioanalytical assay and the integrity of bioanalysis conduct. A conventional ISR involves analysis of at least 20 samples taken from an in vivo study a second time using the method that was described in prestudy validation and employed in generating the initial study sample results. However, this practice is sometimes inadequate to confirm bioanalytical results that are unexpected. The present report discusses several additional exploratory activities that were performed to confirm the unexpected plasma concentration-time results of NVP-1, an investigational drug candidate, observed in the plasma samples collected from patients in a phase II trial. These approaches include (1) LC-MS/MS reanalysis of the study samples after multiple freeze/thaw cycles followed by a short-term benchtop storage, (2) evaluation of additional MS/MS transitions in LC-MS/MS, (3) employment of a different sample preparation procedure in LC-MS/MS, and (4) study sample dilution using plasma samples from healthy volunteers. These procedures are practical and can be readily implemented in the confirmatory LC-MS/MS bioanalysis of other small molecules.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.