Abstract

This paper presents a practical approach to clinical dosimetry for electron beams at extended source-to-surface distance (SSD). Characteristics of electron beams from a Siemens MD-2 accelerator are presented at a nominal SSD of 100 cm and an extended SSD of 115 cm. Relative output factors are measured at both 100 and 115 cm SSDs for a range of square field sizes for each beam energy. The change of output with SSD does not follow the inverse square law, if the nominal SSD is used. This deviation is larger for lower beam energy and smaller field sizes. An effective SSD, , has been determined at for each beam energy as a function of field size. A comprehensive approach to treatment time calculations is proposed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.