Abstract

Increased pressure to obtain more, higher sensitivity data from less sample is especially critical for large peptides, whose already optimized LC-MS methods are heavily challenged by traditional ligand-binding assays. Critical bioanalytical assays were adapted to integrated microscale LC to reduce sample volumes while increasing sensitivity. Assays for teriparatide, glucagon and human insulin and five analogs were transferred from 2.1 mm analytical scale LC to a 150 µm scale system. This resulted in a 15-30 fold overall improvement in sensitivity derived from increased signal to noise, three to six fold reduction in injection volumes, and a two to five fold reduction in sample consumption. Integrated microscale LC reduces sample consumption while enabling single picomolar quantification for therapeutic and endogenous peptides.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.