Abstract
The benefits resulting from the implementation of calibration systems based on the power equation concept are presented. Resulting uncertainties in the comparison of bolometer efficiencies are on the order of ±0.2 percent. Calibration costs have also been cut in half in respect to previous methods practiced at Autonetics. This method also shows significant cost advantages with respect to other methods described in the literature by Dentinger and de Pian. Both manual and automated bolometer calibration systems are described. Results of an experiment designed to illustrate the advantages of this concept using both type-N and SMA connectors are presented.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
More From: IEEE Transactions on Instrumentation and Measurement
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.