Abstract

A practical absolute wavelength meter was assembled separately a reference wavelength source and wavelength measurement system. The frequency (wavelength) stability was obtained to be 2×10-12 for a reference wavelength source. The other hand accuracy of wavelength measurement was estimated to be ** Michelson interferometer with vacuum chamber. The wavelength was estimated to be 1523.48813 nm for a 1523nm He-Ne laser as test laser. Moreover, the pressure dependence for wavelength was measured to be 0.0044 [pm/hPa]. Measurement reproducibility of the wavelength meter was guaranteed to be ±0.035pm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.