Abstract

The potentiostat system consisting of a constant voltage driving negative feedback circuit is very useful in understanding of electrochemical interfacial phenomena. This paper proposes a designing method for a simplified potentiostat circuit for biomedical instrumentations and bio-sensors, by using conventional op-amplifiers and instrumentation amplifiers. The ranges are 10μV∼10mV, 0.1nA∼10mA in the frequency of 10Hz∼10kHz. Results of the loop frequency characteristics, the intrinsic noises and the Lissajous characteristic of a designed potentiostat consisting of low noise and low bias-current op-amps. and instrumentation amps. are shown.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.