Abstract

AbstractCurrent methods for determining the activity of long chain quaternary ammonium compounds (QACS) are based either on dye partition, titration, or colorimetric analysis. The two major disadvantages of these methods are the disparity of partition coefficients among differently constituted QACS and the difficulty in detecting visual end points. Some potentiometric titration methods for QACS have been reported in the literature. However, back titration techniques, as well as complicated electrode systems, are generally involved.A new potentiometric titration system is presented which uses aqueous sodium tetraphenylborate (TPB) solution as a titrant and a platinum‐platinum electrode system to detect the end point. Standard potentiometric titration instruments may be used for this method. This new potentiometric method is superior in precision and accuracy to visual (colorimetric) methods.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.