Abstract

Hydrazinecarbothioamide derivative based ligands 2,2′-(((ethane-1,2-diylbis-(oxy))bis(3-methoxy-4,1-phenylene))bis-(methanylylidene))bis(hydrazinecarbothioamide) (L1), 2,2′-(((propane-1,3-diylbis(oxy))bis(3-methoxy-4,1-phenylene))bis-(methanylylidene))bis-(hydraznecarbothioamide) (L2) and 2,2′-((((oxybis(ethane-2,1-diyl))bis(oxy))bis(3-methoxy-4,1-phenylene))bis(methanylylidene))bis(hydrazinecarbothioamide) (L3) were synthesized and investigated as monitoring probe for Co2+ ion. Fabricated coated graphite electrode (CGE) and coated pyrolytic graphite electrode (CPGE) were prepared and CPGE explored the best performance with optimized membrane composition L2: PVC: 1-CN: KTpClPB ≡ 6:35:56:3 (w/w, mg). CPGE showed low detection limit of 6.1 × 10−9 mol L−1, Nernstian compliance (29.4 ± 0.6 mV decade−1 of activity), wide working pH range (3.0–8.5) and fast response time (7 s). The sensor is practically applicable in real samples and used as an indicator electrode in the potentiometric titration of Co2+ ion with EDTA.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.