Abstract
The potentiometric determination of tetraphenylborate with silver nitrate solution was investigated a) at i=0 in the presence of calomel reference electrode and silver resp. silicone rubber based halide-selective indicator electrodes and b) at i≠0 in the presence of silver (cathode)-calomel, silver-silver, silver-platinum, platinum-silver and graphite-silver electrode couples. An indirect method is described for the determination of potassium ions with potentiometric end-point indication. Experiments were also carried out in order to develop methods for the titration of silver, potassium and thallium(I) ions with sodium tetraphenylborate solution in the presence of anodic polarized graphite and calomel reference electrodes. The relative standard deviations were 0.38–0.49% for tetraphenylborate, 1.96% for potassium and 1.07% for thallium(I).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.