Abstract
Sibutramine (SBT+) ion-selective electrode (ISE) was constructed using poly(vinyl chloride) matrix membrane containing sibutramine–phosphotungstate (SBT–PTA) plasticized with dioctyl phthalate. The electrode exhibits Nernstian response of 59.54 mV/ concentration decade over the concentration range 1.0 × 10−5 to 1.0 × 10−2 M and detection limit of 7.94 × 10−6 M. The electrode was fully characterized in terms of its composition, response time, life span, pH, selectivity, and temperature and was then applied to the potentiometric determination of SBT in its pure bulk and pharmaceutical formulations under batch and flow injection conditions and for the in vitro testing of the dissolution profile of SBT capsules.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.