Abstract
Starting from a warning of possible artificially high plasma colistin concentrations observed in clinical samples in our recently published LC–MS/MS method, we reviewed our analytical assay and demonstrated, for the first time, that caution must be paid to ionization conditions employed in the MS system.The artefact reported here demonstrates that additional experiments should be performed in the course of validation of LC–MS/MS methods to be used in samples containing both colistin and colistimethate.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.