Abstract

A clinical and cadaveric example show the EndoButton (Acufex Microsurgical Inc, Mansfield, MA), used for anterior cruciate ligament endoscopic fixation, flipping outside the extensor mechanism or vastus lateralis rather than flipping directly outside the lateral femoral cortex. This pitfall was caused by overdrilling the femoral socket beyond the recommended 6 mm and overadvancing the EndoButton beyond the required depth to flip the EndoButton. Overdrilling the femoral socket to a depth of 10 mm still allows the EndoButton to rest properly on the cortex without soft tissue interposition. Increasing angles of knee flexion at the time of Endobutton placement decrease the safe distance beyond the lateral femoral cortex for flipping without soft tissue interposition. There is also potential to flip the EndoButton within the substance of the vastus lateralis, but the flipping action is blunted and not discrete. Arthroscopy 1998 Jan-Feb;14(1):66-9

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.