Abstract

An on-line liquid chromatography-immunochemical detection (LC-ICD) system for the quantification of cytokines in cell extracts has been developed using a post-column continuous-flow reaction detection system using fluorescence labelled antibodies. Cytokines eluting from the micro-HPLC column react with antibodies to form fluorescent complexes. In a second step the excess of free antibody is trapped on a cytokine bound support prior to fluorescence detection. The concentration detection limit of the flow injection-ICD system was 50 pM (20 microl injection volume) for interleukin 4 (IL-4). An absolute detection limit of 1 fmol was obtained for IL-4. Similar to ICD systems for small non-protein analytes developed earlier, reaction times were in the order of 1 minute. The immobilised cytokine affinity columns can easily be regenerated and used for months. The present ICD system for interleukins 4, 6, 8 and 10 was coupled to ion exchange-, size exclusion- and reversed phase chromatography. Important parameters (reaction times, reaction conditions) were investigated to get a better understanding of post-column ICD systems for macromolecules.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.