Abstract

The purpose of our study was to assess the potential for radiation dose reduction in digital postinterventional digital mammograms after marker placement. One hundred consecutive cases of marker placement (hook-wire localization or postbiopsy clip marker placement), with 200 full-field digital baseline mammograms (craniocaudal and mediolateral), were included in this prospective trial. For the postinterventional digital mammograms, the milliampere seconds were reduced either by 50% or by 75%. Dose-reduced images were evaluated for sufficient image quality to verify the position of the marker. In 193 of 200 cases (96.5%), image quality was sufficient to verify the correct position of the marker. One (1%) case with insufficient image quality occurred in the 50% dose-reduction group and 6 (6%) in the 75% dose-reduction group (P = 0.06). Our results indicate that under evaluation of each individual case, a dose reduction of 50% to 75% can be recommended in postinterventional digital mammograms.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.