Abstract
BackgroundDissolved oxygen tension (DOT) is an important parameter for evaluating a bioprocess. Conventional means to measure DOT in shake flasks using fixed Clark-type electrodes immersed in the bulk liquid are problematic, because they inherently alter the hydrodynamics of the systems. Other approaches to measure DOT that apply fluorescing sensor spots fixed at the inside wall of a shake flask are also suboptimal. At low filling volumes for cultivating microorganisms with a high oxygen demand, the measured DOT signal may be erroneous. Here, the sensor spot is sometimes exposed to gas in the head space of the flask. Merely repositioning the sensor spot elsewhere in the flask does not address this problem, since there is no location in the shake flask that is always covered by the rotating bulk liquid. Thus, the aim of this prospective study is first, to verify the systemic error of Clark-type electrodes for measuring DOT in shake flasks. The second principle aim is to use the newly built "flexitube optical sensor" to verify potential errors in conventional optical DOT measurements based on fixed sensor spots.ResultsWith the Clark-type electrode, the maximum oxygen transfer capacity in shake flasks rose compared to that of an analogous system without an electrode. This proves changed hydrodynamics in the system with the Clark-type electrode. Furthermore, regarding the sensor spot experiments under oxygen-limited conditions where the DOT value ought to approach zero, the acquired signals were clearly above zero. This implies that the sensor spot is influenced by oxygen present in the headspace and not only by oxygen in the bulk liquid.ConclusionsThe Clark-type electrode is unsuitable for measuring DOT. Moreover, the newly built rotating flexitube optical sensor is useful to verify potential errors of conventional optical DOT measurement techniques applying fixed sensor spots.
Highlights
Dissolved oxygen tension (DOT) is an important parameter for evaluating a bioprocess
The results indicate that a Clark-type electrode acts as a baffle, thereby increasing the whole oxygen transfer of the system
Since Clark-type electrodes act like a baffle, they inherently alter the hydrodynamics of the system and raise the maximum oxygen transfer capacity
Summary
Dissolved oxygen tension (DOT) is an important parameter for evaluating a bioprocess. Conventional means to measure DOT in shake flasks using fixed Clark-type electrodes immersed in the bulk liquid are problematic, because they inherently alter the hydrodynamics of the systems. Other approaches to measure DOT that apply fluorescing sensor spots fixed at the inside wall of a shake flask are suboptimal. Repositioning the sensor spot elsewhere in the flask does not address this problem, since there is no location in the shake flask that is always covered by the rotating bulk liquid. The aim of this prospective study is first, to verify the systemic error of Clark-type electrodes for measuring DOT in shake flasks. The second principle aim is to use the newly built “flexitube optical sensor” to verify potential errors in conventional optical DOT measurements based on fixed sensor spots. Scheidle et al [9] combined an optical pH-measurement and the RAMOS principle in one device
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.