Abstract

To present a case where we noted occult contamination of the G-probe used for transscleral cyclodiode and to critically analyze the techniques used for making a G-probe suitable for repeated use. Close microscopic examination of the G-probes used once for transscleral cyclodiode was done and a critical review of the literature on practice of reuse of the G-probe and the techniques used to "cleanse" the probe before reuse was performed. We noticed microscopic contamination of the G-probe by the tear film fluid in all the probes examined by us. The review of literature indicates that repeated use of the G-probe is not uncommon. The types of techniques used for making it suitable for repeated use indicate that it is not universally recognized that the lumen of the G-probe can accumulate fluid during the procedure, which makes it potentially hazardous when used for other patients. In our opinion the G-probe should not be reused as inadequately reprocessed G-probe can lead to risk of nosocomial infections, serious iatrogenic complications, and medico-legal problems.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.