Abstract

This study sought to investigate the feasibility of using magnetic resonance-magnetic resonance-ultrasound (MR-MR-US) fusion imaging navigation (FIN) with needle tail intelligent positioning (NTIP) to guide puncture in percutaneous transforaminal endoscopic discectomy (PTED). First, in a pig experiment, we found that puncture errors in lumbar intervertebral foramen (LIF) puncture using magnetic resonance-magnetic resonance-ultrasound (MR-MR-US) FIN with NTIP for experienced and novice operators were 2.00 ± 1.00 and 2.57 ± 0.98 mm, respectively (p=0.231), suggesting this technique was minimally dependent on experience. Then, two experienced surgeons agreed (inter-observer agreement к=0.801) that the quality of MR-MR fusion images was good or sufficient. Finally, we performed PTED in eight patients using MR-MR-US FIN with NTIP, and no significant complications were reported during LIF puncture. Overall, MR-MR-US FIN with NTIP may be a potential application for guiding puncture in PTED, but more clinical studies with a larger sample size are required to further evaluate the advantages of MR-MR-US FIN with NTIP.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.