Abstract

PurposeThe placement of localization clips following percutaneous biopsy is a standard practice for a variety of situations. Subsequent clip displacement creates challenges for imaging surveillance and surgical planning, and may cause confusion amongst radiologists and between surgeons and radiologists. Many causes have been attributed for this phenomenon including the commonly accepted “accordion effect.” Herein, we investigate the performance of a low cost surgical clip system against 4 commercially available clips. MethodsWe retrospectively reviewed 2112 patients who underwent stereotactic vacuum-assisted core biopsy followed by clip placement between January 2013 and June 2016. The primary performance parameter compared was displacement >10 mm following vacuum-assisted stereotactic core biopsy. Within the group of clips that had displaced, the magnitude of displacement was compared. ResultsThere was a significant difference in displacement among the clip types (P < .0001) with significant pairwise comparisons between pediatric surgical clips and SecureMark (38% vs 28%; P = .001) and SenoMark (38% vs 27%; P = .0001) in the proportion displaced. The surgical clips showed a significant magnitude of displacement of approximately 25% greater average distance displaced. ConclusionsAs a whole, the commercial clips performed better than the surgical clip after stereotactic vacuum-assisted core biopsy suggesting the surrounding outer component acts to anchor the central clip and minimizes clip displacement. The same should apply to tomosynthesis-guided biopsy.