Abstract

There has been a significant increase in the use of 90Y-microspheres in treating liver malignancies. This increase could be seen over the last 30 y, and Food and Drug Administration approval of 2 products-Sirtex SIR-Spheres and Boston Scientific TheraSphere-has helped in the proliferation of these treatments. As the increase in use of both products rose at our institution, there was a need to determine whether there should be special considerations for patients who receive one product compared with patients who receive the other product. This determination was made by measuring exposure rates for several regions of the patient before and after implantation. An independent-samples t test analysis (ɑ = 0.05) was performed for 50 patients (25 TheraSphere and 25 SIR-Spheres) to determine whether the products behaved similarly to the extent that exposure to others was minimized and that as-low-as-reasonably-achievable principles were kept. The results showed that the products exhibited no significant differences in exposure rates, suggesting that no special considerations are needed for the procedure for one product compared with the other.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.