Abstract

The aim of this study was to describe a quality assurance issue, that is, the production of nondiagnostic high-contrast radiographs when imaging teeth restored with all-zirconia crowns on bitewing radiographs. All-zirconia crowns were imaged with DIGORA Optime photostimulable phosphor (PSP) plates (Soredex/Orion Corp., Helsinki, Finland). To assess the differences in software processing, the PSP plates were scanned into third-party software as well as directly into the twain and native software provided by the manufacturer. Gamma correction, histogram stretch, and scanner resolution settings were adjusted. Vertical bitewings were acquired to increase anatomic coverage. Scanning into third-party software or directly into the twain and native software did not improve contrast. Shifting the lower limit of the histogram stretch to 3 with a gamma correction of 2 resolved the problem. Neither scanner resolution setting nor vertical bitewings improved contrast. The nondiagnostic high-contrast radiographs result from imaging software not effectively displaying the available gray scale. The software processing error appears to be initiated by the high-attenuation characteristics of zirconia. Consequently, radiographs with a high ratio of zirconia crown to normal anatomy are particularly susceptible.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.