Abstract

Conventional bolus calculators apply negative prandial corrections when premeal glucose levels are low. However, no study has evaluated the need for this negative correction with closed-loop systems. We analysed data retrospectively from a cohort study evaluating a closed-loop artificial pancreas system conducted in a diabetes camp over a period of 11 days. Meal boluses with negative correction (n = 98) of 47 participants aged 8 to 22 years were examined. If there was no insulin-on-board from previous boluses at mealtime, the postprandial hyperglycaemia rate increased with increased duration of insulin suspension (P = .03), with odds ratios being exaggerated by 17% per 10 minutes of suspension. However, if there was insulin-on-board from previous boluses, the hyperglycaemia rate did not change with increased duration of insulin suspension (P = .24). When there was no insulin-on-board, the rate of hyperglycaemia after meals preceded by no suspension was 21% (3/14), compared with 52% (12/23) and 64% (9/14) after meals preceded by suspensions of ≥50 and ≥70 minutes, respectively. Meal size did not influence these results. We conclude that, in the absence of insulin-on-board, negative prandial corrections may not be necessary following long insulin suspensions.