Abstract

A 39-year-old man underwent emergency surgery for type A acute aortic dissection complicated by paraplegia. However, hemolytic anemia increased significantly due to severe stenosis of the proximal anastomosis one month after surgery. He finally underwent a redo procedure 4 months after the initial operation whereupon it was verified that half of the inner felt strip used for proximal stump fixation had turned up and was protruding into the inner lumen. We report here on a rare case of survival of postoperative early hemolytic anemia due to severe graft stenosis caused by an inverted inner Teflon felt strip without any extra vascular compression.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.