Abstract

Complications after implantation of 503 plate haptic silicone intraocular lenses were compared with those after implantation of 253 silicone three-piece polypropylene loop lenses. The uncorrected visual acuities were also compared. Capsulorhexis and endocapsular phacoemulsification were used in all cases. The incidence of cystoid macular edema and posterior capsule opacification requiring a YAG capsulotomy were significantly higher in the cases with polypropylene three-piece loop lenses. Pigment dispersion and decentration also occurred more frequently in this group. This is probably because the 10.5 mm plate haptic lens is more rigid than the polypropylene loop lens. It also fixates more posteriorly within the capsular bag than the anteriorly located polypropylene lens. Patients implanted with a plate haptic silicone intraocular lens through a 3.5 mm scleral incision had significantly better uncorrected visual acuity at five to six weeks postoperatively than those with a three-piece silicone lens that required a 4.5 mm incision for insertion.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.