Abstract
Coronary stent fracture has been postulated as an important mechanism for in-stent restenosis and late-stent thrombosis. We have developed a nondestructive microcomputed tomography (micro-CT) technique to image en bloc stented arterial segments. A 43-year-old man died of sepsis 5 months following placement of a paclitaxel-eluting stent (PES) in the left circumflex coronary artery to treat in-stent restenosis of a sirolimus-eluting stent (SES). Micro-CT three-dimensional images of the overlapping stents were notable for numerous connecting element fractures along the length of the SES. No fractures were observed in the PES. The greatest area of luminal narrowing is related to the region of most extensive stent fracture. Micro-CT of en bloc specimens provides high resolution (16 μm), three-dimensional images allowing for visualization of the stented arterial segment in a nondestructive manner.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.