Abstract

BackgroundPulmonary vein isolation (PVI) is less effective in persistent atrial fibrillation (PerAF) than in paroxysmal atrial fibrillation (AF). However, the CAPLA (Effect of Catheter Ablation Using Pulmonary Vein Isolation With vs Without Posterior Left Atrial Wall Isolation on Atrial Arrhythmia Recurrence in Patients With Persistent Atrial Fibrillation: The CAPLA randomized clinical trial) of PVI vs posterior wall isolation (PWI) did not support empiric PWI in PerAF. We examined pulmonary vein (PV) and posterior wall (PW) electrical characteristics to determine if select patients may benefit from additional PWI. ObjectivesThis study sought to determine the impact of PV and PW electrical characteristics on AF ablation outcomes in the CAPLA randomized study. MethodsParticipants in spontaneous AF at the time of ablation were included from the CAPLA study. The mean, shortest, and longest PV, PW, and left atrial (LA) appendage cycle length measurements were annotated preablation using a multipolar catheter for 100 consecutive cycles. Next, cardioversion was performed with a high-density LA voltage map completed. Cox proportional hazards regression was utilized to determine clinical and electroanatomic predictors of AF recurrence overall and according to ablation strategy. Follow-up included twice daily single-lead electrocardiograms or continuous monitoring for 12 months. ResultsA total of 151 patients (27% female, age 65 ± 9 years, 18% long-standing PerAF, LA volume index 52 ± 16 mL/m2, median AF duration 5 months [IQR: 2-10 months]) were in AF on the day of procedure and were randomized to PVI alone (50%) or PVI+PWI (50%) according to the CAPLA randomized clinical trial protocol. Baseline clinical, echocardiographic, and electroanatomic parameters were comparable between groups (all P > 0.05) including PV and PW characteristics. After 12 months, freedom from AF off antiarrhythmic drug therapy was 51.7% in PVI and 49.7% in PVI+PWI (log-rank P = 0.564). Rapid PW activity was defined as less than the median of the shortest PW cycle length (140 ms) and rapid PV activity was defined as less than the median of the shortest PV cycle length (126 ms). In those with rapid PW activity, the addition of PWI was associated with greater arrhythmia-free survival (56.4%) vs PVI alone (38.6%) (HR: 0.78; 95% CI: 0.67-0.94; log-rank P = 0.030). Moreover, in those undergoing PVI only, the risk of AF recurrence was higher in those with rapid PW activity (55.3% vs 46.5% in slower PW activity; HR: 1.50, 95%CI 1.11-2.26; log-rank P = 0.036). Rapid PV activity and PV cycle length (individual PVs or average of all 4 PVs) were not associated with outcome (all P > 0.05) regardless of ablation strategy. There was no correlation between PW cycle length and posterior low voltage (r = −0.06, P = 0.496). The addition of PWI did not improve arrhythmia-free survival in subgroups with LA enlargement (LA volume index >34 mL/m2) (HR: 0.69; 95% CI: 0.39-1.25; P = 0.301), posterior low-voltage zone (HR: 1.06; 95% CI: 0.68-1.66; P = 0.807), or long-standing PerAF (HR: 1.10; 95% CI: 0.71-1.72; P = 0.669). ConclusionsRapid PW activity is associated with an increased risk of AF recurrence post–catheter ablation. The addition of PWI in this subgroup was associated with a significant improvement in freedom from AF compared with PVI alone. The presence of rapid PW activity may identify patients with PerAF likely to benefit from PWI.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call