Abstract

AbstractThe construction and performance of a post‐column reactor for capillary electrophoresis with laser‐induced fluorescence detection in 10 μm i.d. capillaries is described. The post‐column reactor is based on a gap design and relies primarily on diffusion for addition and mixing of derivatizing reagents. Reactors with gaps of 4 to 160 μm are easily built with minimal micromanipulation between two 10 μm i.d. capillaries. Optimal separation efficiencies and sensitivities are obtained with gaps of less than 10 μm. Separations of amino acids and proteins with o‐phthaldialdehyde/2‐mercaptoethanol post‐column derivatization and laser‐induced fluorescence detection are shown. Linear detector response is found for amino acids and proteins, and peak efficiencies as high as 230,000 theoretical plates are obtained for some analytes. The use of relatively small columns for capillary electrophoresis results in a separation and detection system that produces very low mass detection limits for biological molecules from picoliter sample injections. Mass detection limits of 130 and 5.2 attomoles are obtained for glycine and human transferrin, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.