Abstract
The electron cyclotron emission imaging (ECEI) instrument is widely used to study the local electron temperature (Te) fluctuations by measuring the ECE intensity IECE ∝ Te in tokamak plasmas. The ECEI measurement is often processed in a normalized fluctuation quantity against the time averaged value due to complication in absolute calibration. In this paper, the ECEI channels are relatively calibrated using the flat Te assumption of the sawtooth crash or the tearing mode island and a proper extrapolation. The 2-D relatively calibrated electron temperature (Te,rel) images are reconstructed and the displacement amplitude of the magnetohydrodynamic modes can be measured for the accurate quantitative growth analysis.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.