Abstract

A qualitative method using liquid chromatography-atmospheric pressure chemical ionization-mass spectrometry (LC/APCI-MS) has been developed and validated for the identification of trace hexamethylene triperoxide diamine (HMTD) using three structurally-specific ions. Residues are extracted with deionized water (DI) and identified using a gradient mobile phase program and positive ion full scan mode on a Thermo Finnigan LCQ Ion Trap Mass Spectrometer. This method was validated according to several performance characteristics for the qualitative identification of an analyte using the characteristic ions, demonstrating the method’s reliability for use on forensic applications. The method’s limit of detection (LOD) can identify HMTD in an extract from a cotton matrix to which 20μg of HMTD has been applied (equivalent to 10ppm in extract). Previous scientific publications using LC/MS have not demonstrated post-blast HMTD residue analyses and suffer from a lack of chromatographic retention, sufficient number of mass spectral ions with validation, or require more complex/expensive instrumental methods (accurate mass or MS/MS). Post-blast analyses were successfully conducted with two syringe detonations that verified the efficacy of the method on the analysis of debris and residues following detonation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.