Abstract

This work presents an analysis and comparison of the efficacy of two methods for pedicle screw placement during posterior spinal fusion surgery. A total of 100 screws (64 manual and 36 power driven), all placed utilizing a surgical navigation system, were analyzed and compared. Final screw placement was compared to initial surgical plans using the navigation system, and the final screw locations were analyzed on the basis of angular deviation from these planned trajectories as well as screw translation within a critical reference plane. The power driver was found to insignificantly decrease the resulting angular deviation of these pedicle screws with a mean deviation of 3.35 degrees compared to 3.44 degrees with the manual driver (p = 0.853). Conversely, the power driver was found to increase the translational distance in the critical region, with mean deviations of 2.45 mm for the power driver compared to 1.54 mm with the manual driver. The increase in translational deviation was significant (p = 0.002) indicating that there may be some loss in performance from the adoption of the power driver.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.