Abstract

We demonstrate a general technique to achieve a precise radial displacement of the nodal line of the radiofrequency (rf) field in a linear Paul trap. The technique relies on the selective adjustment of the load capacitance of the trap electrodes, achieved through the addition of capacitors to the basic resonant rf circuit used to drive the trap. Displacements of up to ∼100 µm with micrometer precision are measured using a combination of fluorescence images of ion Coulomb crystals and coherent coupling of such crystals to a mode of an optical cavity. The displacements are made without measurable distortion of the shape or structure of the Coulomb crystals, as well as without introducing excess heating commonly associated with the radial displacement of crystals by adjustment through static potentials. We expect this technique to be of importance for future developments of microtrap architectures and ion-based cavity QED.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.