Abstract

To determine whether leg position or distance between the inner electrodes affect the magnitude of impedance change during deep inspiration, 30 adult patients (60 limbs) were studied with the Minnesota Impedance Cardiograph. Phlebography was performed in 19 limbs and correlated with the impedance findings. There was no difference in the results of impedance testing between the straight- and bent-knee positions, nor was there a difference for the two electrode positions. The phlebographic correlation indicated that none of the techniques was accurate enough to warrant use of the impedance method as the sole diagnostic maneuver in deep phlebothrombosis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.