Abstract

PESIC is an integrated front-end for multianode photomultiplier based nuclear imaging devices. Its architecture has been designed to improve position sensitive detectors behavior by equalizing its response over its whole area. Its preamplying stage introduces two main benefits: digitally programmable gain adjustment for every photomultiplier output, and isolation from other front-end electronics by means of current buffers. This last feature allows to use different types of photomultipliers and optimizes front-end deadtime, reducing impact position dependent output delay. PESIC also includes an indirect measurement of the depth of interaction of the gamma ray inside the scintillator crystal, based on the width of its light distribution. Test measurements have been carried out in an experimental dual detector PET setup in order to quantify improvements obtained from integrated front-end calibration capabilities.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.